Slides · 1.2
Figure 1.2. The sensor-selection table. Rows: IEPE, charge, piezoresistive, capacitive, force transducer, strain gauge, laser vibrometer, laser displacement gauge and displacement transducer. Columns: lowest frequency read (zero hertz or a fixed cutoff), temperature (set by the electronics, the element or the converter, with "datasheet" where it varies by model), range (fixed or set at the amplifier), shock (survives, zero shift, damped), what it needs from the rack (constant current, charge amplifier, bridge excitation), and the test it is chosen for.
Sensor Types and What Each Is For
Read · 11 slides
1 / 11
Sensor Types and What Each Is For
- Choose IEPE, charge or piezoresistive for a stated test
- Say when a force gauge, strain gauge or laser instrument is the right channel
The IEPE Accelerometer
- A piezoelectric element with a small amplifier built into the case
- Powered by a constant current down the same coaxial cable that brings the signal back
- One manufacturer's figures: 2 to 20 mA from 18 to 30 V; output on a bias of 8 to 12 V, about 5 V either side
Fixed Range, Fixed Low End
- The range is the full-scale output over the fixed sensitivity
- 100 mV/g reaches 5 V at 50 g; 10 mV/g reaches it at 500 g
- The low-frequency limit is set by a discharge time constant chosen at manufacture
- The temperature limit is the electronics', and varies by model
The Bias Is the First Diagnostic
- Inside its window: the sensor is powered and its amplifier alive
- Near the supply: an open circuit; near zero: a short
- It cannot say whether the sensitivity is right, the sensor mounted, or the axis correct
The Charge Accelerometer
- The same element with no electronics in the case; an external charge amplifier
- Hotter than an IEPE sensor will stand, and a range set for the test
- The cost is the cable: low-noise, short, held still, connectors clean and dry
The Piezoresistive Accelerometer
- A resistor bridge on a flexure: it reads acceleration down to zero hertz
- Needs bridge excitation and an input that accepts the steady part; lower sensitivity
- Method 516.8 requires it where shock response below 2 Hz is wanted
Compression and Shear
- Compression: the element stacked under preload on the base, so base bending and base temperature make output
- Shear: the element loaded sideways, isolated from the base; both effects much smaller
- A bending plate or a moving temperature asks for shear; the datasheet says which design it is
Triaxial Sensors and Transverse Sensitivity
- Three elements, three sensitivities, three rows in the channel table
- The most common fault: an axis assigned to the wrong channel
- Transverse sensitivity: 5 percent or less for 514.8, 7 percent or less for 516.8
- Orientation recorded against a feature of the article and photographed
Other Channels a Test Is Handed
- A force transducer between the fixture and the article: the channel force limiting runs on
- A strain gauge: stretch at a point, closer to what breaks; usually a monitor
- Laser vibrometer, displacement transducer, laser displacement gauge: no mass added
The Sensor-Selection Table
- One row per channel type: lowest frequency, temperature, range, shock, what it needs from the rack, the test it is for
- A printable page for the lab kit
| Type | Lowest frequency read | Temperature limit set by | Range | Shock | Needs from the rack | Chosen for |
|---|---|---|---|---|---|---|
| IEPE | a fixed cutoff, its time constant | the electronics: datasheet | fixed at manufacture | survives its rating; zero shift possible | constant current, 2 to 20 mA | most channels on most tests |
| Charge | a fixed cutoff, set by the chain | the element: far hotter; the converter stays outside the chamber | set at the charge amplifier | survives its rating; zero shift possible | charge amplifier or in-line converter; low-noise cable | the hottest chamber tests |
| Piezoresistive | zero hertz | datasheet | fixed at manufacture | survives; no zero shift; some damped | bridge excitation; DC-coupled input | shock where response below 2 Hz matters |
| Capacitive | zero hertz | datasheet | fixed at manufacture | datasheet | a DC supply; DC-coupled input | long-stroke tests at a few hertz |
| Force transducer | a fixed cutoff (piezoelectric) | datasheet | fixed (IEPE type) or set at the amplifier (charge type) | datasheet; it carries the article's load and preload | constant current or charge amplifier | force limiting: the interface force |
| Strain gauge | zero hertz | the gauge and its adhesive: datasheet | set at the amplifier | datasheet | bridge completion and excitation | stress at a point, as a monitor |
| Laser vibrometer | datasheet | not on the article | set on its own controller | not mounted: nothing to survive | its own controller; a voltage input | light articles; surfaces nobody may touch |
| Laser displacement gauge | zero hertz; its top is a few hundred hertz | not on the article | fixed by its measuring range | not mounted: nothing to survive | its own supply; a voltage input | shock displacement on the calibration load; body motion at the low end |
| Displacement transducer (LVDT or proximity probe) | zero hertz | datasheet | fixed by its stroke | datasheet | its own conditioner and excitation | the shaker's stroke against its rating |
Datasheet marks a value that varies by model: the datasheet for the sensor in hand is the only number that counts. Zero shift is the piezoelectric element's failing after a severe shock.
Decide
- Choose the sensor type from the test, not the drawer: lowest frequency, temperature at the sensor, peak level
- Shock below 2 Hz, or a zero-shift risk: piezoresistive
- Hotter than the IEPE datasheet allows: charge, with its converter outside
- Everything else starts as IEPE; a bending surface or a moving temperature asks for shear